CLIA Waved
The Integrated Drug Test Cup simultaneously detects the presence or absence of:
Methamphetamine (mAMP)
Cocaine (COC)
Marijuana (THC)
3 panel E-Z Split Key Cup II
The cup’s key activates the test by unlocking the specimen from the safety chamber and drops it into the testing chamber.
This drug test is a clinical grade test designed for professional drug testing
No-touch system
USA FDA Cleared
The Integrated E-Z Split Key Cup II is a rapid urine screening test that can be performed without the use of an
instrument. The test utilizes monoclonal antibodies to selectively detect elevated levels of specific drugs in urine.
The Integrated E-Z Split Key Cup II is a lateral flow chromatographic immunoassay for the qualitative
detection of multiple drugs and drug metabolites in urine at the following cut-off concentrations:
Test
Calibrator
Cut-off
Methamphetamine (mAMP)
D-Amphetamine
1,000 ng/mL
Cocaine (COC)
Benzoylecgonine
300 ng/mL
Marijuana (THC)
11-nor-Delta9-THC-9 COOH
50 ng/mL
Directions For Use
Materials Provided
Test card with integrated cup
Package insert
Materials Required But Not Provided
Timer
Analytical Specificity
The following table lists the concentrations of compounds (ng/mL) that are detected as positive in urine by
the Integrated E-Z Split Key Cup II at 5 minutes.
METHAMPHETAMINE
d-Methamphetamine
1000
p-Hydroxymethamphetamine
30000
l-Methamphetamine
8000
3,4-Methylenedioxymethamphetamine (MDMA)
2000
Mephentermine
50000
COCAINE
Benzoylecgonine
300
Cocaine
780
Cocaethylene
12500
Ecgonine
32000
MARIJUANA
11-nor-Delta9 -THC-9 COOH
50
Cannabinol
20000
11-nor-Delta8-THC-9 COOH
30
Delta8 �THC
15000
Delta9 �THC
15000